Skip to main content
Neuroleptics in Serum/Plasma

Neuroleptics in Serum/Plasma LC-MS/MS Analysis Kit

Antipsychotic (AP) drugs are widely prescribed for the treatment of schizophrenia and psychosis. Their effect is based on influencing the synaptic conduction in the central nervous system. They fall into two classes: first-generation or “typical” antipsychotics, and second-generation or “atypical” antipsychotics. The neuroleptics bind to specific receptors of the neurotransmitters such as dopamine, preventing transmission of stimulus information and leading to inhibition of the transmission of neural signals. APs are used in various neuropsychiatric conditions such as attention-deficit hyperactivity disorder (ADHD), behavioural disturbances in dementia, geriatric agitation, depression, eating disorders, personality disorders, insomnia, generalized anxiety disorder, obsessive-compulsive disorder, post-traumatic stress disorder, and substance use and dependence disorders.

  • Highlights

    Simultaneous measurement the concentrations of 26 neuroleptics.

    Total run time 8 min.

    Just a few pipetting steps for the sample treatment-protein precipitate and shoot approach-no need for derivatization.

    Each drug safeguarded by its own internal standard.

    Consuming small volume of patient’s sample.

    Long life span of HPLC column.
  • Parameters

    Aripiprazole, Dehydro-Aripiprazole, Fluphenazine, Flupentixol, Paliperidone (9-Hydroxyrisperidone), Ziprasidone, Risperidone, Zuclopenthixol, Quetiapine, Pipamperone, Haloperidol, Thioridazine, Amisulpride, Sulpiride, Perazine, Zotepine, Levomepromazine, Clozapine, Chlorpromazine, Chlorprothixene, Olanzapine, Norclozapine, Desmethylolanzapine, Norquetiapine, Promethazine and Melperone
  • Matrix

    Serum, plasma

Sample Preparation


  • Step 1
    Pipette 50 µl serum sample/calibrators/control in a centrifuge tube.
  • Step 2
    Add 30 µL of IS solution, vortex for 5 sec.
  • Step 3
    Add 400 µl Reagent-1 to the sample and vortex 5 sec.
  • Step 4
    Centrifuge 3500 rpm for 5 min. and after the centrifuge decant the supernatant into a HPLC vial.
  • Chromatogram

  • METHOD PERFORMANCE

    Analyte LOQ (mg/L)   6460 QQQ Accuracy % 6470 QQQ Linearity R2 6470 QQQ
    Amisulpride  0,001 92,29 0,999
    Aripiprazole  0,001 88,98 0,999
    Chlorpromazine  0,029 97,3 0,999
    Chlorprothixene  0,051 103,71 0,999
    Clozapine 0,107 94,74 0,999
    Dehydro-Aripiprazole  0,001 84,36 0,999
    Desmethylolanzapine  0,227 98,03 0,999
    Flupentixol 0,301 87,61 0,999
    Fluphenazine 0,01 101,84 0,999
    Haloperidol 0 84,18 0,999
    Levomepromazine 0,007 96,19 0,998
    Melperone  0,011 102,33 0,999
    Norclozapine  0,037 95,54 0,999
    Norquetiapine  0,001 90,98 0,998
    Olanzapine  0,002 90,89 0,999
    Paliperidone  0,001 95,39 0,999
    Perazine 0,001 99,61 0,999
    Pipamperone 0,101 94,41 0,999
    Promethazine 0,141 86,05 0,999
    Quetiapine 0,024 89,31 0,999
    Risperidone  0,014 98,46 0,999
    Sulpiride 0,069 100,38 0,999
    Thioridazine 0,004 106,9 0,999
    Ziprasidone 0,002 99,65 0,999
    Zotepine 0,535 102,56 0,993
    Zuclopenthixol 0,28 106,65 0,999
ADAPTING DIAGNOSTICS IN CHROMATOGRAPHY COUPLED MASS SPECTROMETRY..
PRODUCTS
Contact Us

The easiest way is to fill the contact form. Simply mail us to info@jasem.com.tr or give us a call +902165720200