Neuroleptics in Serum/Plasma

Neuroleptics in Serum/Plasma LC-MS/MS Analysis Kit

Antipsychotic (AP) drugs are widely prescribed for the treatment of schizophrenia and psychosis. Their effect is based on influencing the synaptic conduction in the central nervous system. They fall into two classes: first-generation or “typical” antipsychotics, and second-generation or “atypical” antipsychotics. The neuroleptics bind to specific receptors of the neurotransmitters such as dopamine, preventing transmission of stimulus information and leading to inhibition of the transmission of neural signals. APs are used in various neuropsychiatric conditions such as attention-deficit hyperactivity disorder (ADHD), behavioural disturbances in dementia, geriatric agitation, depression, eating disorders, personality disorders, insomnia, generalized anxiety disorder, obsessive-compulsive disorder, post-traumatic stress disorder, and substance use and dependence disorders.

Highlights

Simultaneous measurement the concentrations of 26 neuroleptics.

Total run time 8 min.

Just a few pipetting steps for the sample treatment-protein precipitate and shoot approach-no need for derivatization.

Each drug safeguarded by its own internal standard.

Consuming small volume of patient’s sample.

Long life span of HPLC column.

Parameters

Aripiprazole, Dehydro-Aripiprazole, Fluphenazine, Flupentixol, Paliperidone (9-Hydroxyrisperidone), Ziprasidone, Risperidone, Zuclopenthixol, Quetiapine, Pipamperone, Haloperidol, Thioridazine, Amisulpride, Sulpiride, Perazine, Zotepine, Levomepromazine, Clozapine, Chlorpromazine, Chlorprothixene, Olanzapine, Norclozapine, Desmethylolanzapine, Norquetiapine, Promethazine and Melperone

Matrix

Serum, plasma

Sample Preparation


  • Step 1
    Pipette 50 µl serum sample/calibrators/control in a centrifuge tube.
  • Step 2
    Add 30 µL of IS solution, vortex for 5 sec.
  • Step 3
    Add 400 µl Reagent-1 to the sample and vortex 5 sec.
  • Step 4
    Centrifuge 3500 rpm for 5 min. and after the centrifuge decant the supernatant into a HPLC vial.

Chromatogram

METHOD PERFORMANCE

Analyte LOQ (mg/L)   6460 QQQ Accuracy % 6470 QQQ Linearity R2 6470 QQQ
Amisulpride  0,001 92,29 0,999
Aripiprazole  0,001 88,98 0,999
Chlorpromazine  0,029 97,3 0,999
Chlorprothixene  0,051 103,71 0,999
Clozapine 0,107 94,74 0,999
Dehydro-Aripiprazole  0,001 84,36 0,999
Desmethylolanzapine  0,227 98,03 0,999
Flupentixol 0,301 87,61 0,999
Fluphenazine 0,01 101,84 0,999
Haloperidol 0 84,18 0,999
Levomepromazine 0,007 96,19 0,998
Melperone  0,011 102,33 0,999
Norclozapine  0,037 95,54 0,999
Norquetiapine  0,001 90,98 0,998
Olanzapine  0,002 90,89 0,999
Paliperidone  0,001 95,39 0,999
Perazine 0,001 99,61 0,999
Pipamperone 0,101 94,41 0,999
Promethazine 0,141 86,05 0,999
Quetiapine 0,024 89,31 0,999
Risperidone  0,014 98,46 0,999
Sulpiride 0,069 100,38 0,999
Thioridazine 0,004 106,9 0,999
Ziprasidone 0,002 99,65 0,999
Zotepine 0,535 102,56 0,993
Zuclopenthixol 0,28 106,65 0,999